Ząbek, Andrzej, Małecka, Barbara, Nowosielska-Ząbek, Ewa, Dębski, Maciej ORCID: https://orcid.org/0000-0002-3669-3916, Boczar, Krzysztof and Lelakowski, Jacek (2019) Switch between AAI and DDD mode pacing-What is the mechanism? Annals of Noninvasive Electrocardiology, 24. ISSN 1542-474X
Full text not available from this repository. (Request a copy)Abstract
The electrocardiogram (ECG) interpretation in patients with implantable pacemaker is often a perplexing problem. The difficulty in the device evaluation increases in the presence of novel timing cycles and additional functions. Authors describe a special function frequently encountered in Medtronic dual-chamber pacemakers and implantable cardioverter-defibrillator devices called managed ventricular pacing (MVP) and demonstrate its performance in the patient with undersensing episodes in ventricular channel. Intermittent ventricular undersensing in the device with MVP feature turned on caused repetitive mode switches between AAI and DDD mode. This report shows unexceptional occurrence of tricky ECG findings in patient with Medtronic dual-chamber device.
Item Type: | Article |
---|---|
Faculty \ School: | Faculty of Medicine and Health Sciences > Norwich Medical School |
Related URLs: | |
Depositing User: | LivePure Connector |
Date Deposited: | 24 Jan 2023 18:30 |
Last Modified: | 24 Apr 2023 14:30 |
URI: | https://ueaeprints.uea.ac.uk/id/eprint/90778 |
DOI: | 10.1111/anec.12648 |
Actions (login required)
View Item |