Evaluation of four HbA1c point-of-care devices using international quality targets: Are they fit for the purpose?

Lenters-Westra, Erna and English, Emma ORCID: https://orcid.org/0000-0002-0480-6744 (2018) Evaluation of four HbA1c point-of-care devices using international quality targets: Are they fit for the purpose? Journal of Diabetes Science and Technology, 12 (4). pp. 762-770. ISSN 1932-2968

[thumbnail of Accepted manuscript]
Preview
PDF (Accepted manuscript) - Accepted Version
Download (1MB) | Preview

Abstract

BACKGROUND: Point-of-care (POC) testing is becoming increasingly valuable in health care delivery, and it is important that the devices used meet the same quality criteria as main laboratory analyzers. While external quality assessment (EQA) provides a great tool for assessing quality, many POC devices are not enrolled in these schemes and standard laboratory evaluations are needed to assess performance. METHODS: The Clinical and Laboratory Standards Institute (CLSI) protocols EP-5 and EP-9 were applied to investigate imprecision, accuracy and bias. We assessed bias using the mean of 4 certified secondary reference measurement procedures (SRMPs). RESULTS: The Afinion2™ and the Quo-Lab had CVs of ≤1.7 and ≤2.4% respectively in IFCC SI units (≤1.2 and ≤1.7% NGSP) and a bias ≤2 mmol/mol (≤0.2% NGSP) at 48 and 75 mmol/mol (6.5 and 9.0% NGSP). Sigma for the Afinion2 was 5.8 and for the Quo-Lab 4.0. Both methods passed the NGSP criteria with 2 instruments when compared with 4 individual SRMPs. The HbA1c 501 had a CV of 3.4% and 2.7% in IFCC SI units (2.1% and 1.7% NGSP) and a bias ≤2.4 mmol/mol (≤0.2% NGSP) and passed the NGSP criteria with 2 instruments compared with 4 individual SRMPs except for instrument 2 compared with the Tosoh G8. Sigma was 2.1. The A1Care had a sigma of 1.4 and failed all criteria mainly due to a high CV (6.2% and 4.1% in IFCC SI units [4.1% and 2.9% NGSP] at 48 and 75 mmol/mol [6.5 and 9.0% NGSP]). CONCLUSIONS: The analytical performance was excellent for the Afinion2 and the Quo-Lab, acceptable for the HbA1c 501 and unacceptable for the A1Care according to different used criteria, demonstrating that whilst performance is improving there are still areas for considerable improvement.

Item Type: Article
Uncontrolled Keywords: diabetes,hba1c,hb-variants,point of care,sigma metrics,sdg 3 - good health and well-being ,/dk/atira/pure/sustainabledevelopmentgoals/good_health_and_well_being
Faculty \ School: Faculty of Medicine and Health Sciences > School of Health Sciences
UEA Research Groups: Faculty of Medicine and Health Sciences > Research Groups > Health Promotion
Depositing User: LivePure Connector
Date Deposited: 13 Sep 2018 13:30
Last Modified: 25 Sep 2024 13:33
URI: https://ueaeprints.uea.ac.uk/id/eprint/68276
DOI: 10.1177/1932296818785612

Downloads

Downloads per month over past year

Actions (login required)

View Item View Item