Tofts, P. S. and Johnson, G. (1987) A SAFE (saturate after the echo) sequence to eliminate an artefact in the measurement of T2 in partially saturated NMR systems. Physics in Medicine and Biology, 32 (10). pp. 1345-1353. ISSN 0031-9155

Full text not available from this repository. (Request a copy)

Abstract

Presents a modified spin-echo pulse sequence which eliminates the artefact described. The sequence will work under almost any conditions of echo time and repetition time. The behaviour of the sequence has been verified experimentally. The magnitude of the error caused by not using a SAFE pulse and assuming that the naive equation given is correct has been calculated in detail for a range of experimental parameters.

Item Type: Article
Uncontrolled Keywords: chlorides, manganese compounds,manganese,humans,magnetic resonance spectroscopy,diagnostic use,instrumentation,methods
Faculty \ School: Faculty of Medicine and Health Sciences > Norwich Medical School
UEA Research Groups: Faculty of Medicine and Health Sciences > Research Groups > Norwich Clinical Trials Unit
Faculty of Medicine and Health Sciences > Research Groups > Cancer Studies
Depositing User: Pure Connector
Date Deposited: 07 Oct 2015 15:00
Last Modified: 15 Dec 2022 02:42
URI: https://ueaeprints.uea.ac.uk/id/eprint/54601
DOI: 10.1088/0031-9155/32/10/013

Actions (login required)

View Item View Item